First check for wave guide connections and the setup in general.
Check water cooling is ON.
Open the helium line. If the line is not being used for long time, please
purge the line.
Open the helium cylinder nob, open the control nob to pressure of about
1 bar.
Open the valve before the needle valve.
Open the needle valve slowly, and monitor the pressure. The FG1 in the VUV
PCU should reach ~1 ✕ 10-1 mbar. Roughly two full rotation
in the needle valve, there is a blue mark.
Turn on the electronics of the helium lamp. Wait for few minutes.
If there is any red light/sign on the interlock, press the reset switch. It
should clear all warnings/red lights.
Switch ON, and immediately OFF the HV ON/OFF switch to turn on the filament
only. Wait for 1 minute.
Then switch ON the HV ON/OFF switch, and look in the viewport if there is
peach color light (a color between orange and yellow). If not switch off the
high voltage, and adjust the Helium pressure.
Check the reflection from wave guide. It should be close to zero. Otherwise,
adjust the helium pressure.